Command Reference
R&S
®
ZNC
682
User Manual 1173.9557.02 ─ 13
Setting parameters:
<UnitRef>
COMPlex | MAGNitude | PHASe | REAL | IMAGinary | SWR |
GDELay | L | C
Keyword for the physical unit of the response values. The param-
eters form four groups: COMPlex, REAL, IMAGinary, and SWR
select dimensionless numbers (U) for the limit line. MAGNitude
selects relative units (dB). PHASe selects phase units (deg). GDE-
Lay selects time units (s). L selects inductance units (H/Henry). C
selects capacitance units (F/Farad).
Usage:
Setting only
CALCulate<Chn>:LIMit:RDOMain:SPACing
<Format>
Deletes the existing limit line and (re-)defines the physical units of the response values
of the limit line. The units of the stimulus values are defined via
.
Tip:
This command is complemented by
Suffix:
<Chn>
.
Channel number used to identify the active trace.
Setting parameters:
<Format>
LINear | LOGarithmic | DB | SIC
Keyword for the physical unit of the response values.
Default unit: dB (irrespective of the parameter selected)
Usage:
Setting only
CALCulate<Chn>:LIMit:LOWer:STATe
<Boolean>
CALCulate<Chn>:LIMit:UPPer:STATe
<Boolean>
These commands switch the lower and upper limit check on or off. Lower limit line seg-
ments are assigned even numbers; upper limit line segments are assigned odd numbers;
see
.
CALCulate<Chn>:LIMit:LOWer:STATe
does not affect segments
with odd numbers;
CALCulate<Chn>:LIMit:UPPer:STATe
does not affect segments
with even numbers.
Note:
to switch on or off the entire limit check,
including upper and lower limit lines.
Suffix:
<Chn>
.
Channel number used to identify the active trace.
Parameters:
<Boolean>
ON | OFF - switch limit check on or off.
*RST:
OFF
R&S ZVR/ZVAB Compatible Commands