Command Reference
R&S
®
ZNC
393
User Manual 1173.9557.02 ─ 13
CALCulate<Chn>:LIMit:DELete:ALL
Deletes all limit line segments.
Suffix:
<Chn>
.
Channel number used to identify the active trace.
Example:
*RST; :CALC:LIM:CONT 1 GHZ, 1.5 GHZ
Define an upper limit line segment in the stimulus range between
1 GHz and 1.5 GHz, using default response values.
CALC:LIM:DATA 1,1500000000, 2000000000,2,3
Define an upper limit line segment in the stimulus range between
1.5 GHz and 2 GHz, assigning response values of +2 dB and +3
dB.
CALC:LIM:DEL:ALL
Delete both created limit line segments.
Usage:
Event
Manual operation:
See
"Add / Insert / Delete / Delete All"
CALCulate<Chn>:LIMit:DISPlay[:STATe]
<Boolean>
Displays or hides the entire limit line (including all segments) associated to the active
trace.
Suffix:
<Chn>
.
Channel number used to identify the active trace.
Parameters:
<Boolean>
ON | OFF - Limit line on or off.
*RST:
OFF
Example:
*RST; :CALC:LIM:CONT 1 GHZ, 2 GHZ
Define an upper limit line segment in the stimulus range between
1 GHz and 2 GHz, using default response values.
CALC:LIM:DISP ON
Show the limit line segment in the active diagram.
Manual operation:
See
CALCulate<Chn>:LIMit:FAIL?
Returns a 0 or 1 to indicate whether or not the limit check has failed. 0 represents pass
and 1 represents fail
Tip:
Use
to perform a composite (global) limit check.
Suffix:
<Chn>
.
Channel number used to identify the active trace.
SCPI Command Reference