GUI Reference
R&S
®
ZNC
183
User Manual 1173.9557.02 ─ 13
The limit line segment is calculated as a straight line connecting the two points (<Start
Stimulus>, <Start Response>) and (<Stop Stimulus>, <Stop Response>); see
ter 3.4.1.1, "Rules for Limit Line Definition"
Remote command:
Add / Insert / Delete / Delete All
The first four buttons below the segment list extend or shorten the list. The analyzer
places no restriction on the number of segments in a limit line.
●
Add
adds a new segment to the end of the list. The new segment extends from the
"Stop Stimulus" value of the last segment to the end of the sweep range. Its response
values are equal to the "Stop Response" value of the last segment.
●
Insert
adds a new segment before the active segment (marked by a blue background
in the first column of the segment list). The new segment extends from the "Stop
Stimulus" value of the segment before the active segment to the "Start Stimulus"
value of the active segment. Its response values are equal to the "Start Response"
value of the active segment. The segment numbers in the list are adapted.
If no segment is active, "Insert" is equivalent to "Add".
●
Delete
removes the selected segment from the list.
●
Delete All
clears the entire segment list so it is possible to define or load a new limit
line.
Remote command:
Recall / Save Limit Line...
The buttons open an Open File / Save File dialog to load a limit line from a limit line file
or store the current limit line configuration to a file.
Limit line files are ASCII files with the default extension
*.limit
and a special file format.
See
chapter 3.4.1.4, "File Format for Limit Lines"
Remote command:
Trace Settings