Command Reference
R&S
®
ZNC
391
User Manual 1173.9557.02 ─ 13
Note:
The generated segments are upper or lower limit line segments, depending on the
setting.
CALCulate<Ch>:LIMit:CONTrol[:DATA]
does not overwrite the type setting.
Tip:
To define additional new limit line segments without overwriting the old segments
.
Suffix:
<Chn>
.
Channel number used to identify the active trace.
Parameters:
<FreqPowTime>
Default unit: Hz
<FreqPowTime>
Pair(s) of frequency, power or time values, to be defined in accord-
ance with the sweep type (
chapter 3.4.1.1, "Rules for Limit Line Definition"
Range:
Virtually no restriction for limit segments.
*RST:
A segment that is created implicitly, e.g. by means of
CALCulate<Ch>:LIMit:UPPer[:DATA] or CALCu-
late<Ch>:LIMit:LOWer[:DATA,], covers the maxi-
mum sweep range of the analyzer.
Default unit: Hz
Example:
*RST; :CALC:LIM:CONT 1 GHZ, 2 GHZ
Select a lin. frequency sweep (default) and define an upper limit
line segment in the stimulus range between 1 GHz and 2 GHz,
using default response values (
–
40 dB).
CALC:LIM:DISP ON
Show the limit line segment in the active diagram.
Manual operation:
See
CALCulate<Chn>:LIMit:CONTrol:SHIFt
<LimShift>
Shifts an existing existing limit line in horizontal direction. See also
Suffix:
<Chn>
.
Channel number used to identify the active trace.
Setting parameters:
<LimShift>
Offset value for the limit line
Range:
Virtually no restriction for limit segments
Default unit: Hz
SCPI Command Reference