Command Reference
R&S
®
ZNC
392
User Manual 1173.9557.02 ─ 13
Example:
*RST; :CALC:LIM:CONT 1 GHZ, 2 GHZ
Define a limit line segment in the stimulus range between 1 GHz
and 2 GHz, using default response values.
CALC:LIM:CONT:SHIF 1; :CALC:LIM:CONT?
Shift the segment by 1 Hz. The modified limit line segment ranges
from
1000000001
(Hz) to
2000000001
(Hz).
Usage:
Setting only
Manual operation:
See
CALCulate<Chn>:LIMit:DATA
<Type>, <StartStim>, <StopStim>, <StartResp>,
<StopResp>
Defines the limit line type, the stimulus and response values for a limit line with an arbitrary
number of limit line segments. See
chapter 3.4.1.1, "Rules for Limit Line Definition"
Note:
In contrast to
, this command does
not overwrite existing limit line segments. The defined segments are appended to the
segment list as new segments.
Suffix:
<Chn>
.
Channel number used to identify the active trace.
Parameters:
<Type>
Identifier for the type of the limit line segment:
0
–
limit line segment off, segment defined but no limit check per-
formed.
1
–
upper limit line segment
2
–
lower limit line segment
Range:
0, 1, 2 (see above)
<StartStim>,
<StopStim>,
<StartResp>,
<StopResp>
Stimulus and response values of the first and last points of the limit
line segment.
Range:
Virtually no restriction for limit segments
Default unit: The default stimulus units depend on the sweep type:
Hz for frequency sweeps, dBm for power sweeps, s
for time sweeps. The default response units are the
units of the selected trace format (CALCu-
late<Chn>:FORMat).
Example:
*RST; :CALC:LIM:CONT 1 GHZ, 1.5 GHZ
Define an upper limit line segment in the stimulus range between
1 GHz and 1.5 GHz, using default response values.
CALC:LIM:DATA 1, 1500000000, 2000000000, 2, 3
Define an upper limit line segment in the stimulus range between
1.5 GHz and 2 GHz, assigning response values of +2 dB and +3
dB.
CALC:LIM:DISP ON
Show the limit line segment in the active diagram.
Manual operation:
See
SCPI Command Reference