Command Reference
R&S
®
ZNC
394
User Manual 1173.9557.02 ─ 13
Example:
*RST; :CALC:LIM:CONT 1 GHZ, 2 GHZ
Define an upper limit line segment in the stimulus range between
1 GHz and 2 GHz, using default response values.
CALC:LIM:STAT ON; FAIL?
Switch the limit check on and query the result.
CALC:LIM:STAT:AREA LEFT, TOP
For a subsequent check at the GUI or a hardcopy, move the pass/
fail message to the top left position.
Usage:
Query only
Manual operation:
See
CALCulate<Chn>:LIMit:SEGMent<Seg>:AMPLitude:STARt
<Response>
CALCulate<Chn>:LIMit:SEGMent<Seg>:AMPLitude:STOP
<Response>
These commands change the start or the stop response values (i.e. the response values
assigned to the start or stop stimulus values) of a limit line segment. A segment must be
created first to enable the commands (e.g
). See also
chapter 3.4.1.1, "Rules for Limit Line Definition"
Tip:
To define the response values of several limit line segments with a single command,
.
Suffix:
<Chn>
.
Channel number used to identify the active trace.
<Seg>
Segment number
Parameters:
<Response>
Response value
Range:
Virtually no restriction for limit segments
*RST:
The default response values of a segment that is cre-
ated by defining its stimulus values only (e.g. by
means of CALCulate<Ch>:LIMit:CONTrol[:DATA]),
are -40 dB.
Default unit: dB
Example:
CALC:LIM:DATA 1,1500000000, 2000000000,2,3
Define an upper limit line segment (segment no. 1) in the stimulus
range between 1.5 GHz and 2 GHz, assigning response values of
+2 dB and +3 dB.
:CALC:LIM:SEGM:AMPL:STAR 5; STOP 5; :CALC:LIM:
SEGM:TYPE LMIN
Change the segment to a lower limit line segment with a constant
response value of +5 dB.
CALC:LIM:DATA?
Query the type, the stimulus and response values of the created
segment with a single command. The response is
2,1500000000,2000000000,5,5.
Manual operation:
See
SCPI Command Reference