Command Reference
R&S
®
ZNC
399
User Manual 1173.9557.02 ─ 13
Parameters:
<Response>
Default unit: dBm
<Response>
Pair(s) of response values
Range:
Virtually no restriction for limit segments
*RST:
The response value of a segment that is created by
defining its stimulus values only (e.g. by means of
CALCulate<Chn>:LIMit:CONTrol[:DATA]), is -40 dB.
Default unit: dB
Example:
CALC:LIM:LOW -10, 0, 0, -10
Define two lower limit line segments covering the entire sweep
range. Two upper limit line segments with default response values
are created in addition.
CALC:LIM:UPP -10, 0, 0, -10
Change the response values of the upper limit line segments.
CALC:LIM:DISP ON
Show the limit line segments in the active diagram.
Manual operation:
See
CALCulate<Chn>:LIMit:LOWer:FEED
<StimulusOffset>, <ResponseOffset>[,
<TraceName>]
CALCulate<Chn>:LIMit:UPPer:FEED
<StimulusOffset>, <ResponseOffset>[,
<TraceName>]
Generates a lower or an upper limit line using the stimulus values of a data or memory
trace and specified offset values.
Suffix:
<Chn>
.
Channel number used to identify the active trace. This trace pro-
vides the stimulus data for the limit line unless another trace
<TraceName>
is specified.
Setting parameters:
<StimulusOffset>
Stimulus offset value, used to shift all imported limit line segments
in horizontal direction.
Range:
-1000 GHz to +1000 GHz [Hz]
*RST:
0 Hz
Default unit: Hz
<ResponseOffset>
Response offset value, used to shift all imported limit line seg-
ments in vertical direction.
Range:
-1012 dB to +1012 dB
*RST:
0 dB
Default unit: dB
<TraceName>
Name of the selected trace as used e.g. in
. If no trace name is specified the analyzer
uses the active trace no.
<Chn>
.
SCPI Command Reference