GUI Reference
R&S
®
ZNC
245
User Manual 1173.9557.02 ─ 13
Use "Import Cal Kit..." to add new kits to the list.
Remote command:
Use Sliding Match
Available for cal kits which contain a Sliding Match standard; see
If selected, the Sliding Match appears in the list of measured standards whenever the
selected calibration type requires a Match. For a valid calibration, either the Match or at
least three positions of the Sliding Match must be measured.
Remote command:
n/a
Same Connector / Gender All Ports
Assigns the same connector type or gender to all selected physical ports. For some multi-
port calibration types, the port connector types must be equal, e.g. because they require
a Through standard with known characteristics.
Remote command:
Import Cal Kit
Opens the Import Calibration Kit dialog to load and (if desired) activate a cal kit file. See
chapter 3.5.2.4, "Cal Kit Files"
By default cal kit files are stored in the
C:\Users\Public\Documents\Rohde-Schwarz\Vna\Calibration
directory.
●
Three different
import
file formats are supported: R&S ZVA-specific binary cal kit
files (
*.calkit
), ZVR-specific binary cal kit files (
*.ck
), cal kit files in Agilent-spe-
cific ASCII formats (
*.csv
,
*.prn
.
Remote command:
Back
Go back to
.
Start
Start
.
Step 3: Calibration
Allows to acquire error correction data for every required port (pair) and calibration stand-
ard, where "required" depends on the selected ports and calibration type. On "Apply" the
R&S ZNC calculates the system error correction (error terms) from the measurement
data of the standards and applies the result to the active channel.
Access:
CHANNEL > CAL > Start Cal > Start... (Manual) > Next > Next
Channel Settings