Command Reference
R&S
®
ZNC
396
User Manual 1173.9557.02 ─ 13
Example:
CALC:LIM:DATA 1,1500000000, 2000000000,2,3
Define an upper limit line segment (segment no. 1) in the stimulus
range between 1.5 GHz and 2 GHz, assigning response values of
+2 dB and +3 dB.
CALC:LIM:SEGM:STIM:STAR 1GHZ; STOP 2 GHZ; :
CALC:LIM:SEGM:TYPE LMIN
Change the segment to a lower limit line segment with a stimulus
range between 1 GHz and 2 GHz.
CALC:LIM:DATA?
Query the type, the stimulus and response values of the created
segment with a single command. The response is
2,1000000000,2000000000,2,3.
Manual operation:
See
CALCulate<Chn>:LIMit:SEGMent<Seg>:TYPE
<LimLineType>
Selects the limit line type for a limit line segment. This can be done before or after defining
the stimulus and response values of the segment, however, a segment must be created
first to enable this command (e.g
CALC:LIM:DATA
).
Note:
The type command overwrites the
settings and
is overwritten by them. It is not affected by the other commands in the
CALCulate<Chn>:LIMit...
subsystem defining stimulus and response values of limit
lines.
Suffix:
<Chn>
.
Channel number used to identify the active trace.
<Seg>
Segment number
Parameters:
<LimLineType>
LMIN | LMAX | OFF
Limit line type
Range:
LMAX (upper limit line segment), LMIN (lower limit
line segment), OFF (limit check switched off, limit line
segment not deleted)
*RST:
LMAX
Example:
*RST; :CALC:LIM:UPP 0, 0
Define an upper limit line segment across the entire sweep range,
using a constant upper limit of 0 dBm.
CALC:LIM:SEGM:TYPE LMIN
Turn the defined limit line segment into a lower limit line segment.
Manual operation:
See
CALCulate<Chn>:LIMit:SOUNd[:STATe]
<Boolean>
Switches the acoustic signal (fail beep) on or off. The fail beep is generated each time
the analyzer detects an exceeded limit.