Calibration Guide
CKW Indicators
Ohaus Corporation www.ohaus.com
i
CKW Indicators
Calibration Guide
PREPARATION
CKW Indicators have two calibration processes:
Span
and
Linearity
.
Ensure that the scale is level, stable, and undisturbed during calibration.
(Otherwise the process could fail.)
Allow Indicator to warm up for five minutes after stabilizing to room temperature.
Use suitable calibration masses (available from Ohaus).
Exact weight value depends on model. Follow prompts for indicated values to use.
(Weight values here are examples only.)
Set the scale’s calibration units to match the calibration masses: in the Setup Menu,
choose either kilograms or pounds.
(See Instruction Manual for details.)
If the scale is set to Legal for Trade operation, a seal must be broken to unlock it.
(See the Instruction Manual.)
Note
: To cancel calibration, press
or turn the scale off.
SPAN CALIBRATION
5
(Release button)
mMeNU
mMeNU
mMeNU
mMeNU
CCCCAL
AL
AL
AL
1
1
(Flashing: clear platform)
Spa
Spa
Spa
SpaNNNN
0000
1
(Acquiring zero value.)
(Place weight on pan.)
*
----CCCC----
30
30
30
30
1
(Establishing calibration weight.)
*
For a different value, press
No
.
Enter new value via the numeric keypad.
Place weight on platform. Press
Yes
.
----CCCC----
Scale returns to weighing mode.
Visit www.balances.com to see the full line of Ohaus Balances
Buy your next Ohaus Balance online at https://scaleman.com/