background image

Calibration Guide 

 

CKW Indicators

 

Ohaus Corporation  www.ohaus.com 

i

 

CKW Indicators 

Calibration Guide 

PREPARATION 

CKW Indicators have two calibration processes: 

Span

 and 

Linearity

  Ensure that the scale is level, stable, and undisturbed during calibration. 

(Otherwise the process could fail.) 

  Allow Indicator to warm up for five minutes after stabilizing to room temperature. 

  Use suitable calibration masses (available from Ohaus). 

Exact weight value depends on model. Follow prompts for indicated values to use. 

(Weight values here are examples only.)

 

  Set the scale’s calibration units to match the calibration masses: in the Setup Menu, 

choose either kilograms or pounds. 

(See Instruction Manual for details.)

 

 

If the scale is set to Legal for Trade operation, a seal must be broken to unlock it.  

(See the Instruction Manual.) 

Note

: To cancel calibration, press 

 

or turn the scale off. 

 

SPAN CALIBRATION 

5

 

  

(Release button) 

 

 

mMeNU

mMeNU

mMeNU

mMeNU    

 

CCCCAL

AL

AL

AL

 

 

1

 

 

1

 

(Flashing: clear platform)

 

 

Spa

Spa

Spa

SpaNNNN

 

 

0000

 

 

1

 

  

(Acquiring zero value.)

    

 

(Place weight on pan.)

*

 

 

 

----CCCC----

 

 

30

30

30

30    

 

1

 

(Establishing calibration weight.)

 

 

For a different value, press 

No

.  

Enter new value via the numeric keypad.  
Place weight on platform. Press 

Yes

.  

 

----CCCC----

 

Scale returns to weighing mode. 

 

Visit www.balances.com to see the full line of Ohaus Balances

Buy your next Ohaus Balance online at https://scaleman.com/

Отзывы: