Command Reference
R&S
®
ZNC
556
User Manual 1173.9557.02 ─ 13
[SENSe:]CORRection:COLLect:AUTO:CKIT:PORTs
<Characterization>,
<TestPort1>, <CalUnitPort1>, <TestPort2>...
Generates a characterization (cal kit file) with the specified name containing the cal kit
data of the active calibration unit (
). The cal kit file can be stored in the calibration unit or written to a directory on the ana-
lyzer.
, but with flexible port
assignment.
Setting parameters:
<Characterization>
Name and (possibly) directory of the created characterization / cal
kit file (string parameter, extension *.calkit): If no path and directory
is specified, the file is stored in the calibration unit (see also
[SENSe<Ch>:]CORRection:COLLect:AUTO
). The factory cal-
ibration data on the unit is not overwritten. If a path and directory
is specified, the file is stored in the directory on the analyzer's hard
disk. The default directory (
MMEMory:CDIRectory
) is not used.
<TestPort1>
Number of first test port.
<CalUnitPort1>
Number of the calibration unit port to whom the first test port
(<TestPort1>) is assigned.
<TestPort2>
Number of second test port...
Usage:
Setting only
Manual operation:
See
[SENSe:]CORRection:COLLect:AUTO:CKIT:PORTs:ADD
<Characterization>,
<TestPort1>, <CalUnitPort1>, <TestPort2>...
Extends an
existing
characterization of the active calibration unit (
) with additional, previously uncharacterized
cal unit ports.
This functionality is not available at the GUI.
Setting parameters:
<Characterization>
Name and (possibly) directory of an existing characterization (e.g.
created using
<TestPort1>
Number of first test port.
<CalUnitPort1>
Number of the (previously uncharacterized!) calibration unit port,
to whom the first additional test port (<TestPort1>) is assigned.
<TestPort2>
Number of the second test port ...
SCPI Command Reference