Command Reference
R&S
®
ZNC
650
User Manual 1173.9557.02 ─ 13
Parameters:
<GBIBAddress>
GPIB address (integer number)
Range:
0 to 30
Example:
SYST:COMM:GPIB:ADDR 10
Set the GPIB address to 10.
*RST; :SYST:COMM:GPIB:ADDR?
After a reset, the address is maintained (the response is 10).
Manual operation:
See
SYSTem:COMMunicate:GPIB[:SELF]:RTERminator
<Terminator>
Sets the receive terminator of the analyzer. The receive terminator indicates the end of
a command or a data block.
The receive terminator setting is relevant if block data is transferred to the analyzer (
REAL
). In the default setting LFEoi, the analyzer recognizes an LF
character sequence with or without the EOI control bus message as a receive terminator.
An accidental LF in a data block can be recognized as a terminator and cause an inter-
ruption of the data transfer.
The EOI setting is especially important if commands are transferred in block data format,
because it ensures that the parser for command decoding is activated by the terminator
only after the command has been completely transferred. Readout of binary data does
not require a change of the receive terminator.
Parameters:
<Terminator>
LFEoi | EOI
LF character sequence with or without EOI recognized.
Example:
SYST:COMM:GPIB:RTER EOI
Set the terminator to EOI.
SYSTem:COMMunicate:RDEVice:AKAL:ADDRess
<USBAddress>
Selects one of the USB-connected calibration units for calibration (see commands
SENSe<Ch>:CORRection:COLLect:AUTO...
). This command is not necessary if
only one cal unit is connected.
Parameters:
<USBAddress>
Name (USB address) of a connected calibration unit (string vari-
able). The names of all connected cal units can be queried using
SCPI Command Reference