CM-2018 Digital AC/DC Clamp Meter Manual 

This product is a precision instrument, be sure to read this "Instruction Manual" carefully before 
use, so as not to cause damage and injury to you due to misuse. 

1 .   Summary 

The meter is a stable performance, high reliability 3 5/6 digital AC/DC clamp meter, the meter 
adopts 14mm word height LCD display, clear readings. 
It can be used to measure DC voltage, AC voltage, DC current, AC current, variable frequency 
voltage and current, LoZ (low impedance input) false voltage test, resistance, capacitance, 
frequency, temperature, diode, continuity test and non-contact AC voltage detection. It is designed 
with clamp jaw lightning function, backlight display, unit symbol display, data hold, 
automatic/manual range conversion, automatic power off and alarm function. The whole machine 
adopts a microprocessor capable of directly driving LCD and double-integral A/D conversion IC, a 
digital display driver providing high resolution and high accuracy. The meter is fully functional, 
highly accurate in measurement and easy to use, and is an ideal tool for laboratories, factories, radio 
enthusiasts and families. 

2. Security guidelines   

The meter complies with IEC1010-1 (EN61010-1) pollution2, CAT.III 600V and UL3111-1 
standards. Please read this manual carefully before use. 

2-1. Electrical symbols 

  "   "CAUTION: Refer to instructions before use. "

  "Low voltage symbol. "   "WARNING! 

Danger of electric shock." "Double insulation. 

2-2. Safety instructions and precautions for use 

2-2-1. Read this manual before use, and operate in accordance with the requirements specified in the 
manual to avoid accidents that endanger personal safety and damage the instrument! 
2-2-2. Before opening the case, please disconnect the input signal and remove the test lead, in order 
to avoid electric shock or damage to the meter, no water should leak into the case. 
2-2-3. Do not use this meter for test work when the housing is not installed or its fixing screws are 
not tightened. 
2-2-4. Do not input the value higher than the limit at every range during measurement. 
2-2-5. In the resistance range terminal, it is prohibited to access the voltage value. 
2-2-6. Do not use the LPF low-pass filter option to verify the presence of dangerous voltages, which 
may be present in excess of the indicated value. The voltage must be measured without the filter 
connected to detect the presence of dangerous voltages. 
2-2-7. Do not use LoZ mode to measure voltages in circuits that may be damaged by the low 
impedance (300 kΩ) of this mode. 
2-2-8. The power switch should be set to OFF after use. 
2-2-9. For long-term storage, please remove the battery to avoid damage to the internal components 
by battery leakage. 
2-2-10. A voltage higher than 60V DC or 30V AC RMS will produce a serious risk of electric shock. 
2-2-11. Be especially careful when clamping uninsulated conductors or busbars, accidental contact 
with conductors can cause electric shock. 

3.Integrated specification 
3-1. General Characteristics

 

3-1-1. display mode: LCD liquid crystal display.   
3-1-2. Maximum display: 6000 (3 5/6) digit automatic polarity display.   
3-1-3. Maximum jaw opening: 28mm. 
3-1-4. Over-range display: "OL" is displayed in the highest position. 
3-1-5. Hold: Data hold. 
3-1-6. Relative value measurement. 
3-1-7. Sampling rate: 3 times/second. 
3-1-8. Battery low voltage display: 

"

" symbol appears. 

3-1-9. Continuity test: buzzer sounds when it is less than 50Ω. 
3-1-10. Automatic range or manual range. 
3-1-11. Automatic power off. 
3-1-12. Power consumption: about 3mA. 
3-1-13. Power source: two 1.5V batteries ("AAA" 7# batteries). 
3-1-14. Working environment: (0

40)℃, relative humidity

70%RH. 

3-1-15. External dimensions: 185mm×65mm×32mm(L×W×H). 
3-1-16. Weight: about 200g (including battery). 
3-1-17. Accessories: user manual, a certificate of conformity

test leads, temperature sensor TP01 

banana probe, handbag, gift box, two 1.5V batteries. 

3-2. Technical specifications 

3-2-1. Accuracy: ± (a% r word count), guaranteed for 1 year. ambient temperature (23±5) 
°C, relative humidity <70%. 
3-2-2. Temperature coefficient: When the temperature is <18℃ or >28℃, the additional 
temperature coefficient error is 0.1×the specified accuracy/℃. 
3-2-3. DC Voltage 

Measurement range 

Accuracy 

Resolving power 

600mV 

± (1.0% r 5) 

0.1mV 

6V 

± (0.5% of r 5) 

0.001V 

60V 

0.01V 

600V 

0.1V 

LoZ 0V-600V 

± (2.0% r 10) 

0.1V 

Input impedance: mV range > 40MΩ. other ranges input impedance is 10MΩ. LoZ V is 300kΩ. mV 
range open circuit will have unstable digital display can be stable after connected to the load < ±3 
words). 
Maximum input voltage ±600V, 

610V with alarm sound. LCD display "OL" when input >610V. 

 

3-2-4. AC Voltage (ACV True RMS) 

Measurement range 

Accuracy 

Resolving power 

6V 

± (0.8% of r 5) 

0.001V 

60V 

0.01V 

600V 

0.1V 

LPF 0V-600V 

± (4.0% r 5) 

0.1V 

LoZ 0V-600V 

± (2.0% r 10) 

0.1V 

Input impedance: 10MΩ, LoZ V is 300kΩ. Accuracy guarantee range: 5~100% range, short circuit 
allows less than 10 words of remaining reading. 
Sine wave true RMS value is displayed. Frequency response: 40Hz~1kHz for the 600V range, 
40Hz~2kHz for the rest of the range. 
Maximum input voltage 600Vrms, 

610V with alarm sound. LCD display "OL" when input 

>610V. 
 
3-2-5. DC Current 

Measurement range 

Accuracy 

Resolving power 

60A 

± (1.5% of r 5) 

10mA 

600A 

100mA 

800A 

1A 

Overload protection: 1000A (input time not exceeding 60 seconds) 
Note: Clear to zero before measurement. The measured current conductor should be placed through 
the middle of the clamp jaw.     
 
3-2-6. AC Current (True RMS) 

Measurement range 

Accuracy 

Resolving power 

60A 

± (1.5% of r5) 

10mA 

600A 

100mA 

800A 

1A 

Frequency response: 40Hz- 1kHz for sine wave and triangle wave. 40Hz-200Hz for others.   
Overload protection: 1000A (input time not exceeding 60 seconds) 
Note: The measured current conductor should be placed through the middle of the clamp jaw.   
3-2-7. Resistance (Ω) 

Measurement range 

Accuracy 

Resolving power 

600Ω 

± (0.8% of r5)   

0.1Ω 

6 k Ω 

 

± (0.8% of r1) 

1Ω 

60kΩ 

10Ω 

600kΩ 

100Ω 

6 MΩ 

1kΩ 

60MΩ 

± (1.0% of r5) 

10kΩ 

Open circuit voltage: 500mV. 
Overload protection: 250 V rms. 
Warning: Input of voltage values in this range is prohibited! 
 
3-2-8. Capacitance (C) 

Measurement range 

Accuracy 

Resolving power 

6nF 

± (5.0% of r10) 

1pF 

60nF 

10pF 

600nF 

± (2.5% of r5) 

100pF 

6uF 

1nF 

60uF 

10nF 

600uF 

± (5.0% of r10) 

100nF 

6mF 

1uF 

60mF 

10uF 

Overload protection: 250V rms. 
Warning: Input of voltage values in this range is prohibited! 
 
3-2-9. Frequency: (F) 

Measurement range 

Accuracy 

Resolving power 

10Hz~10MHz 

± (0.5% of r4) 

0.001Hz~1kHz 

Input sensitivity: >0.7V. Overload protection: 250V rms 
 
3-2-10. Temperature   

Measurement range 

Accuracy 

Resolving power 

(-40~1000) °C 

<400°C: ± (1.0% of r5) 

400°C: ± (1.5% of r15) 

1 ℃ 

(-40~1832) ℉ 

<750°F: ± (1.0% of r 5) 

750°F: ± (1. 5% of r 15) 

1°F 

Sensor: K-type banana plug (NiCr-NiSi). 
Warning: Input of voltage values in this range is prohibited! 
 
 

3-2-11. Diode – Continuity Test 

Measurement 

range   

Display value   

Test conditions 

 

 

Diode forward voltage drop. 

Forward DC current 0.8mA, open circuit 

voltage about 2.2V. 

The buzzer sounds long when the resistance 

value of two points is less than 50Ω. 

Open circuit voltage is about 2V, press 

"SELECT/LPF" function to switch. 

Overload protection: 250V rms. 
Warning: Input of voltage values in this range is prohibited! 

4. OPERATION 
4-1. Panel Description

 

1

Clamp: 0A to 800A DC AC current and non-contact voltage detecting. 

2

Plier head trigger: pulling the plier head trigger can open the plier head. 

3

Hand protection: a safety design to protect the user's hands from 

touching the dangerous area. 

4

Clamp light: open the clamp lightning function to illuminate the measured area in the dark to 

avoid danger.   

5

NCV indicator: detect the existence of the high voltage in the surrounding area to prevent the 

risk of electric shock. 

6

Knob switch: Used to switch and select the function and range, turn on/off the meter. 

Function 

Description 

NCV 

Non-contact AC voltage detection. 
 
 

 

DC current and AC current measurement, press "SELECT/LPF" to switch the 
measurement. 

 

AC/DC voltage measurement. Long press "SELECT/LPF" key to switch low-pass 
filter measurement. Press Hz/DUTY to toggle frequency/duty cycle measurement 

to 

accommodate high frequency and voltage amplitude measurement.

 

LoZ 

 

Low impedance voltage measurement to determine if there is a spurious voltage in the 
circuit. 
 
 

 

Press "SELECT/LPF" to select resistance measurement, diode, continuity 
measurement, and capacitance measurement cycle switching. 

Hz % 

Frequency measurement, press Hz/DUTY key to switch the frequency/duty cycle 
measurement, 

adapt to frequency voltage amplitude below 10V signal 

measurement.

 

℃/℉ 

For temperature measurement, press the "SELECT/LPF" key to switch between °C 
or °F. 

7

Function Keys 

SELECT/LPF keys. 

1) This is key fo selecting function, base on a working principle of trigger action. Short press this 

key to choose measurement modes: choose DC or AC in    state. choose the Ω /

  /   /    

in            state. choose °C or °F in temperature measurement state. switch frequency/duty cycle 
measurement in AC voltage measurement state.   
2) In ACV/ACA mode, press SELECT/LPF to enter LPF low-pass filter. 

An internal specific filtering circuit to filter out high-
frequency beyond 1kHz interference signals to ensure the accuracy. It can be applied to the 
occasions with variable frequency voltage and current. 

3) The meter will be auto power off when there is no operation within 15minutes, then it will enter 

the sleeping mode, the buzzer will sound 5 times of alarm reminder within 1 minute before auto 
power off, please press this key to turn on the meter if you need to restart the device. 

4) Press this key and put it on Hold and then turn on the Power to awake the meter from sleeping 

mode, and then auto power off function will be canceled.     

REL/ RANGE keys. 

1) In 

  and 

  measurement is "REL" function, press this key to clear the reading to zero and 

enter the relative value measurement, the display appears "REL" symbol, press again to exit the 
relative value measurement. (In 

  and 

  measurement, the display does not return to zero 

before measurement, you need to press this key to make the display to "0" before measurement.) 

2) In the voltage and resistance measurement is "RANGE" function, press this key to select the 

automatic range or manual range working mode. The instrument starts in the auto range state, 
displaying the "AUTO" symbol, press this function to change to manual range, press once to 
increase one gear, cycle from low to high. Press this key continuously for more than 2 seconds to 
return to the automatic range state.

 

HOLD/   key. 

It is a key for Data Hold function to keep the display reading, based on a working principle of 
trigger action, short press this key to lock the display value on 
screen, kept unchanged, and then press other keys HOLD 
function is canceled. Press this key for more than 2 seconds to 
turn on the display Backlight function, press this key again for 
2 seconds to turn off the backlight, if you do not press this key 
after opening it will automatically go out after 30 seconds. 

Hz/DUTY/

  key. 

1) When measuring AC voltage (current), follow this function 

to switch the frequency/duty cycle/voltage (current), and 
switch the frequency/duty cycle (1-99%) when measuring 
frequency. 

2) Lighting control key, press for more than 2 seconds lighting on, and then press for more than 2 

seconds, the lighting will turn off.

 