GUI Reference
R&S
®
ZNC
285
User Manual 1173.9557.02 ─ 13
●
"Sweep" means that each trigger event starts an entire sweep, according to the cur-
rent sweep configuration.
●
"Point" means that each trigger event starts the measurement at the next sweep point.
●
"Partial Measurement" means that each trigger event starts the next partial mea-
surement at the current or at the next sweep point. If each measurement points
requires only one partial measurement, this option is equivalent to "Point".
See also
chapter 3.1.4.1, "Partial Measurements and Driving Mode"
●
"Segment" means that each trigger event starts the next sweep segment within the
current sweep. If a sweep type other than "Segmented Sweep" is active, this option
is equivalent to "Sweep".
Relation with other sweep settings
Some sweep settings are logically incompatible with a particular selection of the triggered
measurement sequence:
●
If a "Time" sweep is performed, the sequence is always a sweep.
●
"Alternate" sweep mode only makes sense if the triggered measurement sequence
comprises more than one sweep point. If "Point" or "Partial Measurement" is selected,
Alternate sweep mode is switched off and vice versa.
Note:
The period of the trigger events should be adjusted to the triggered measurement
sequence. If the analyzer receives a trigger event while the last sequence is still running,
the R&S ZNC skips the trigger event and generates a message.
Remote command:
Delay
Specifies a delay time between the trigger event and the start of the measurement.
The Delay time entered must be zero or positive, so that the trigger event precedes the
start of the measurement (post-trigger).
In "Multiple Trigger" mode the trigger delay can be selected individually using the "Trigger
Manager" dialog. See
chapter 4.4.5.2, "Trigger Manager (Dialog)"
Remote command:
Slope / Level
Specifies the "External" trigger mode in detail.
●
Rising Slope / Falling Slope
means that the rising or falling slope of every external
trigger pulse can trigger a single measurement sequence.