GUI Reference
R&S
®
ZNC
286
User Manual 1173.9557.02 ─ 13
●
High Level / Low Level
means that the analyzer performs a free run measurement
as long as the external trigger signal is high / low. The measurement is discontinued
when the trigger signal changes to low / high.
Remote command:
Manual Trigger
Generates the trigger event for "Manual" trigger mode (see
Free Run / External / Manual /
). "Manual Trigger" is disabled unless manual trigger mode is active.
Remote command:
MANual
*TRG
4.4.5.2
Trigger Manager (Dialog)
The "Trigger Manager" dialog defines individual trigger sources and delays for the trig-
gered measurement sequences. The settings are valid for "Multiple Trigger" mode. For
detailed specifications of the trigger signals refer to
Background information
Refer to
chapter 3.1.4.1, "Partial Measurements and Driving Mode"
Access:
CHANNEL > TRIGGER > Trigger > Trigger Manager...
Fig. 4-4: Example of a multiple trigger configuration
The table in the "Trigger Manager" dialog contains several editable (white) or non-editable
(gray) columns. All settings are analogous to the general trigger settings in the CHANNEL
> TRIGGER > Trigger tab. Refer to the following sections:
●
●
"Free Run / External / Manual / Multiple Triggers"
●
●
Channel Settings