Command Reference
R&S
®
ZNC
568
User Manual 1173.9557.02 ─ 13
Manual operation:
See
[SENSe<Ch>:]CORRection:COLLect:METHod:DEFine
<CalName>, <CalType>,
<TestPort1>, <TestPort2>...
Selects a calibration type for channel
<Ch>
at arbitrary test ports.
Suffix:
<Ch>
.
Channel number of the calibrated channel
Parameters:
<CalName>
Name of the calibration (string parameter). The name serves as a
reference to delete a particular set of system correction data (
<CalType>
REFL | RSHort | FOPort | FTRans | RTRans | FRTRans | OPTPort |
TOSM | UOSM | TRL | TOM | TSM | TRM | TNA | PTOSm | PTOM |
PTSM | PTRM | PTRL | PTNA | PUOSm
Calibration types; see list below.
<TestPort1>
First calibrated test port number. For an n-port calibration type, n
port numbers must be specified. If more than n numbers are
defined, the spare numbers (the last ones in the list) are ignored.
Entering less than n numbers causes an error message. For a one
path two port calibration (
OPTPort
), the first port no. denotes the
node port (fully calibrated port).
<TestPort2>
Second calibrated port number.
Example:
See
Manual operation:
See
The supported calibration types are listed below.
Parameter
Type ("Calibration Presetting" dialog)
REFL | RSHort
Refl Norm Open | Refl Norm Short
FOPort
Full One Port ("Refl OSM")
FTRans | RTRans | FRTRans
Trans Norm - Forward | Trans Norm - Reverse | Trans Norm Both
"Forward" ("Reverse") means that the port with the smaller (larger)
port number serves a a drive port.
OPTPort
One Path Two Ports
TOSM
TOSM
UOSM
UOSM
TRL | TOM | TSM | TRM | TNA
TRL | TOM | TSM | TRM | TNA
PTOSm | PUOSm | PTRL | PTOM |
PTSM | PTRM | PTNA
One of the previous calibration types in combination with a power
calibration (for SMARTerCal).
SCPI Command Reference