Command Reference
R&S
®
ZNC
558
User Manual 1173.9557.02 ─ 13
Example:
CORR:COLL:AUTO:PORT '', 1, 2, 2, 4, 4, 1
Perform an automatic 3-port calibration at the analyzer test ports
1, 2, and 4 using the calibration unit's default calibration kit file and
ports 2, 4, and 1 of the cal unit.
Usage:
Setting only
Manual operation:
See
[SENSe:]CORRection:COLLect:AUTO:PORTs:CONNection?
Returns the assignment between the network analyzer ports and the ports of the con-
nected automatic calibration unit.
Example:
CORR:COLL:AUTO:PORT '', 1, 2, 2, 4, 4, 1
Perform an automatic 3-port calibration at the analyzer test ports
1, 2, and 4 using the calibration unit's default calibration kit file and
ports 2, 4, and 1 of the cal unit.
CORR:COLL:AUTO:PORT:CONN?
Query the actual port assignment. If the cal unit is properly con-
nected according to the previous command, the response is
1,2,2,4,3,0,4,1. A zero means that the corresponding analyzer port
is not connected to any port of the calibration unit.
Usage:
Query only
Manual operation:
See
[SENSe<Ch>:]CORRection:COLLect:AUTO:PORTs:TYPE
<CalType>[,
<Characterization>, <TestPort1>, <CalUnitPort1>, <TestPort2>...]
Selects and initiates an automatic calibration at arbitrary test and calibration unit ports.
A progress monitor for the calibration sweeps is displayed.
Tip:
This command is necessary if the analyzer fails to detect the cal unit ports connected
to each of its ports (e.g. because of a high attenuation in the test setup). If auto-detection
works you can use the simpler command
.
Suffix:
<Ch>
.
Channel number of the calibrated channel
Setting parameters:
<CalType>
FNPort | FOPort | OPTPort | FRTRans | FTRans | RTRans |
UTRans | REFL | RSHort | PFNPort
FNPort - full n-port (UOSM) calibration
FOPort - full one-port ("Refl OSM") calibration
OPTPort - one path two ports calibration
SFTPort - separate full two-port calibration (UOSM calibration at
two ports)
PFNPort
–
for future extensions
SCPI Command Reference