Command Reference
R&S
®
ZNC
381
User Manual 1173.9557.02 ─ 13
Manual operation:
See
6.3.1.4
CALCulate:FILTer[:GATE]...
The
CALCulate:FILTer[:GATE]...
commands define the properties of the time gate
which is used to optimize the time domain response.
........................................................................381
................................................................381
........................................................382
.................................................................382
.................................................................383
...................................................................383
.................................................................383
..................................................................384
...................................................................384
...............................................................384
.................................................................385
CALCulate:FILTer[:GATE]:TIME:AOFFset
<Boolean>
Activates the operating mode where the time gate is moved in the opposite direction when
the "Delay" setting is changed.
Parameters:
<Boolean>
ON | OFF - enable or disable "Adjust Time Gate".
*RST:
OFF
Example:
*RST; :CALCulate1:TRANsform:TIME:STATe ON
CALCulate1:FILTer:GATE:TIME:STATe ON; SHOW ON
Activate time domain representation and a time gate in channel
no. 1. Display the time gate
CALCulate1:FILTer:GATE:TIME:STARt 2ns; STOP 3
ns
Restrict the time gate to the time interval between 2 ns and 3 ns.
CALCulate:FILTer:GATE:TIME:AOFFset ON
Activate an offset of the time gate according to a new delay setting.
SENSe1:CORRection:EDELay1:TIME 1ns
Specify a 1 ns delay at port 1.
CALCulate1:FILTer:GATE:TIME:STARt?; STOP?
Query the time gate position. The response is
1E-009;2E-009
.
Manual operation:
See
CALCulate<Chn>:FILTer[:GATE]:TIME:CENTer
<CenterTime>
Defines the center time of the time gate.
Suffix:
<Chn>
.
Channel number used to identify the active trace.
SCPI Command Reference