Command Reference
R&S
®
ZNC
555
User Manual 1173.9557.02 ─ 13
Usage:
Setting only
Manual operation:
See
[SENSe:]CORRection:COLLect:AUTO:CKIT:PASSword
<Password>
Enters a password to enable a single password-protected action in the automatic cali-
bration or in the characterization wizard. If password protection has been activated man-
ually in the "Characterize Cal Unit" dialog, the password is required for any single exe-
cution of one the following commands, provided that a cal kit (characterization) file dif-
ferent from the active characterization is specified:
[SENSe]:CORRection:COLLect:AUTO
[SENSe]:CORRection:COLLect:AUTO:CKIT
[SENSe]:CORRection:COLLect:AUTO:PORTs
[SENSe]:CORRection:COLLect:AUTO:PORTs:TYPE
[SENSe]:CORRection:COLLect:AUTO:TYPE
Tip:
You have to send
[SENSe:]CORRection:COLLect:AUTO:CKIT:PASSword
repeatedly if your command script uses several password-protected commands. You do
not need a password to perform automatic calibrations using the active cal unit charac-
terization.
Setting parameters:
<Password>
Password (string parameter), as defined in manual control.
Example:
SENSe:CORRection:COLLect:AUTO:CKIT:PASSword
'My_password'
Enter a password
My_password
(assuming that password pro-
tection has been activated manually).
SENSe:CORRection:COLLect:AUTO '', 1, 2
Perform an automatic 2-port calibration at test ports 1 and 2 using
the calibration unit's default calibration kit file and automatic port
assignment.
SENSe:CORRection:COLLect:AUTO:CKIT:PASSword
'My_password'
Re-enter the password.
SENSe:CORRection:COLLect:AUTO '', 1, 2
Repeat the calibration.
Usage:
Setting only
Manual operation:
See
"Authentication / Set Password"
SCPI Command Reference