Command Reference
R&S
®
ZNC
502
User Manual 1173.9557.02 ─ 13
Manual operation:
See
INITiate<Ch>[:IMMediate]:SCOPe
<Scope>
Selects the scope of the single sweep sequence. The setting is applied in single sweep
mode only (
OFF
).
This command is needed in the "ZVR" and "ZVABT" compatibility modes only (
'ZVR' | 'ZVABT'
).
Suffix:
<Ch>
.
Channel number. The channel is ignored because the command
affects all channels.
Parameters:
<Scope>
ALL | SINGle
in all channels or in the referenced channel
<Ch>
only.
Example:
See
Manual operation:
See
6.3.9 INSTrument Commands
The
INSTrument...
commands select or query particular resources (SCPI: logical
instruments) of the analyzer.
............................................................................................502
...........................................................................................503
INSTrument:NSELect
<Channel>
Selects a channel as the active channel.
Parameters:
<Channel>
Number of the channel to be activated. The channel must be cre-
ated before using
ON
.
Range:
1, 2, ...
*RST:
1
Example:
CONF:CHAN2:STAT ON; :INST:NSEL?
Create channel no. 2 and select it as the active channel. The query
returns 2.
INSTrument:PORT:COUNt?
Returns the number of test ports (Port 1, Port 2 ...) of the analyzer.
SCPI Command Reference