Command Reference
R&S
®
ZNC
663
User Manual 1173.9557.02 ─ 13
Return values:
<Frequency>
MINimum | MAXimum
Return minimum or maximum frequency.
Example:
See
Usage:
Query only
Manual operation:
See
"Start / Stop / Center / Span Frequency (Power / Time)"
SYSTem:IDENtify:FACTory
Resets the response to the
*IDN?
query to the factory default value. This command
overwrites a user-defined identification string; see example.
Example:
SYSTem:IDENtify:STRing 'MyDevice'; *IDN?
Define an identity string. The response is
MyDevice
.
SYSTem:IDENtify:FACTory; *IDN?
Re-activate the factory setting. The response is
Rohde-Schwarz,<instrument type>-<ports>,<stock
no.><serial no.>,<firmware version>
.
Usage:
Event
Manual operation:
See
SYSTem:IDENtify[:STRing]
<InstId>
Defines an identity string for the network analyzer. The query is equivalent to
*IDN?
.
Parameters:
<InstId>
String parameter containing the instrument identity
Example:
See
Manual operation:
See
SYSTem:KLOCk
<Boolean>
Locks or unlocks the local controls of the analyzer. This includes the front panel keys,
the keyboard, or other local interfaces.
Parameters:
<Boolean>
ON | OFF - lock or unlock the local keys.
*RST:
OFF
Example:
SYST:KLOCk ON
Lock the local keys.
SYSTem:LANGuage
<Language>
Specifies the remote language for the analyzer.
SCPI Command Reference