Command Reference
R&S
®
ZNC
501
User Manual 1173.9557.02 ─ 13
Parameters:
<Boolean>
ON - the analyzer measures continuously in all channel, repeating
the current sweep. The query returns ON (1) if at least one channel
is measured continuously.
OFF - the measurement is stopped after the number of sweeps
defined via
initiates a new measurement cycle.
*RST:
ON
Example:
See
Manual operation:
See
"All Channels Continuous / on Hold"
INITiate[:IMMediate]:ALL
Starts a new single sweep sequence in all channels. This command is available in single
sweep mode only (
OFF
). The data of the last sweep (or
of previous sweeps, see
Retrieving the Results of Previous Sweeps
) can be read using
SDATa, <count>
.
This command is not supported in the "ZVR" or "ZVABT" compatibility modes.
Note:
In contrast to all other commands of the analyzer, the
INITiate<Ch>[:IMMediate]...
commands have been implemented for overlapped
execution; see
chapter 5.4, "Command Processing"
Example:
See
Usage:
Event
Manual operation:
See
INITiate<Ch>[:IMMediate][:DUMMy]
Starts a new single sweep sequence. This command is available in single sweep mode
only (
OFF
). The data of the last sweep (or previous
sweeps, see
chapter 7.2.4.3, "Retrieving the Results of Previous Sweeps"
on page 726) can be read using
SDATa,
<count>
.
Note:
In contrast to all other commands of the analyzer, the
INITiate<Ch>[:IMMediate]...
commands have been implemented for overlapped
execution; see
chapter 5.4, "Command Processing"
Suffix:
<Ch>
.
Channel number. If the channel number does not exist the ana-
lyzer returns an error message. If the "ZVR" or "ZVABT" compat-
ibility mode is active and
ALL
is selected, this suffix is ignored.
Example:
See
Usage:
Event
SCPI Command Reference