Command Reference
R&S
®
ZNC
500
User Manual 1173.9557.02 ─ 13
Parameters:
<Boolean>
ON - the analyzer measures continuously, repeating the current
sweep.
OFF - the measurement is stopped after the number of sweeps
defined via
initiates a new measurement cycle.
*RST:
ON
Example:
*RST; :CALC2:PAR:SDEF 'TRC2','S11'
DISPlay:WINDow:TRACe2:FEED 'Trc2'
Reset the analyzer to create the default channel no. 1 and default
trace. Create a second trace and display the trace. Both traces are
measured continuously.
INIT1:CONT OFF
Activate single sweep mode for the first channel. The measure-
ment in channel no. 1 is stopped after the current sweep. Channel
no. 2 is still measured continuously.
INIT1
Activate a new (single) sweep in channel no. 1. Channel no. 2 is
still measured continuously.
INIT:CONT:ALL OFF
Activate single sweep mode for all channels. The measurement in
channel no. 2 is also stopped after the current sweep.
INIT:ALL
Re-start a single sweep in both channels.
Example:
Alternative settings using the R&S ZVAB compatibility mode:
*RST; :SYSTem:LANGuage 'ZVABT'
*RST; :INIT:CONT OFF
Activate single sweep mode for all channels (including channel no.
2 created later).
INIT:SCOP SING
State that a single sweep will be performed in the active channel.
CALC2:PAR:SDEF 'TRC2','S11'; :INIT2
Create channel no. 2 with a new trace and start a single sweep in
channel no. 2. Activate single sweep mode for all channels, start
a single sweep in the second channel.
Manual operation:
See
INITiate:CONTinuous:ALL
<Boolean>
Qualifies whether the analyzer measures in single sweep or in continuous sweep mode.
This command is not supported in the "ZVR" or "ZVABT" compatibility modes.
SCPI Command Reference