background image

http://www.omega.com   

e-mail: [email protected]

®

User’s Guide

DP3001

DIGITAL PANEL VOLTMETER

ROHS 2 COMPLIANT

Содержание DP3001

Страница 1: ...http www omega com e mail info omega com User sGuide DP3001 DIGITAL PANEL VOLTMETER ROHS 2 COMPLIANT...

Страница 2: ......

Страница 3: ...nversion 2 2 7 Display 2 2 8 Power 2 2 9 Environmental 2 2 10 Mechanical 3 3 0 MECHANICAL ASSEMBLY INSTALLATION 3 3 1 Case Dimensions 3 3 2 Panel Mounting 4 4 0 POWER SIGNAL 5 4 1 Changing Operating V...

Страница 4: ...1 Side View of Transformer 6 Figure 4 2 Rear Terminal Hookups 7 Figure 5 1 Display Board Jumper Locations 10 Figure 5 2 Main Board Jumper Locations 11 Figure 5 3 Solder Switch Locations 12 Figure 7 1...

Страница 5: ...incorporated in the mains supply cord Furthermore to provide protection against excessive energy being drawn from the mains supply in case of a fault in the equipment an over current protection device...

Страница 6: ......

Страница 7: ...del contains the required circuitry for full range selection and easy configuration 2 0 SPECIFICATIONS 2 1 ANALOG INPUT Model 3 4 5 Range 1 9999 Vdc 19 999 Vdc 199 99 Vdc Resolution 0 1 mV 1 mV 10 mV...

Страница 8: ...ogrammable internally or at the rear connector Overrange indication 4 least significant digits flash 2 8 POWER Standard AC power voltage 115 or 230 Vac 15 AC frequency range 49 to 440 Hz Optional DC p...

Страница 9: ...ouble row of 18 pins 3 96 mm 0 156 in spacing between pins Barrier terminal strip with six 6 screw Bezel Depth behind bezel with connector Panel cutout Weight Case material D1 connector option D4 conn...

Страница 10: ...the panel cutout refer to case dimensions shown below 3 Push the slide retainers onto the case to fit tightly against the panel rear Rotate slide clamps back into original position and tighten clamp s...

Страница 11: ...with the shield terminated to analog ground at the connector Both analog and digital ground are internally connected and should not be connected externally For proper grounding connect the low signal...

Страница 12: ...d Green Green Model Power Signal AC D1 Connector A High 17 High 2 Low 16 Low C Earth Ground T Analog Ground D4 Connector 1 High 6 High 2 Low 5 Low 3 Earth Ground 4 Analog Ground DC D1 Connector 2 DC 1...

Страница 13: ...E24 H 199 99 DP3 7 Spare E23 J 19 999 DP2 8 Spare E22 K 1 9999 DP1 9 Spare E25 L Decimal point select 10 Spare E26 M Spare E27 11 4 6 Vdc power output N Spare E29 12 4 7 Vdc power output P Excitation...

Страница 14: ...s If taken low the converter will complete the full measurement cycle and then hold this reading as long as HOLD is low A positive pulse greater than 300 ns will now initiate a new measurement cycle b...

Страница 15: ...lder switches as indicated Remove all push on jumpers not used 5 1 DECIMAL POINT SELECTION Figure 5 1 Display Board Jumper Locations Decimal Point S1 Alternate decimal point configuration using main b...

Страница 16: ...ons Refer to Figure 5 3 S3 A is used for signal connections Refer to Section 4 for more information Input Configuration Solder Switches Push on Jumpers Wire Jumper Open Close S2 S4 W10 2 Vdc J L H I K...

Страница 17: ...the stability and internal accuracy If calibration is needed use the following procedure 1 Remove the front lens Insert a blade screwdriver under the notch at the bottom of the lens and gently pry it...

Страница 18: ...12 7 0 DRAWINGS Figure 7 1 Assembly Diagram Main and Display Board...

Страница 19: ...t the slope in volts sec during this second period T2 is constant and independent of input signal levels The time required to discharge the capacitor back to zero volts is then proportional to the sig...

Страница 20: ..._______________________ _____________________________________________________________________________ ______________________________________________________________________________ ___________________...

Страница 21: ..._________________________ _____________________________________________________________________________ ______________________________________________________________________________ _________________...

Страница 22: ..._________________________ _____________________________________________________________________________ ______________________________________________________________________________ _________________...

Страница 23: ...NS Equipment sold by OMEGA is not intended to be used nor shall it be used 1 as a Basic Component under 10 CFR 21 NRC used in or with any nuclear installation or activity or 2 in medical applications...

Страница 24: ...Indicators R Turbine Paddlewheel Systems R Totalizers Batch Controllers pH CONDUCTIVITY R pH Electrodes Testers Accessories R Benchtop Laboratory Meters R Controllers Calibrators Simulators Pumps R In...

Отзывы: