SAFETY CONSIDERATIONS

This device is marked with the international Caution symbol.   It is important to read 

this manual before installing or commissioning this device as it contains important 

information relating to Safety and EMC (Electromagnetic Compatibility).

Unpacking & Inspection

Unpack the instrument and inspect for obvious shipping damage. Do not attempt to 

operate the unit if damage is found.

 This instrument is a panel mount device protected in accordance with Class I of EN 

61010 (115/230 AC power connections). Installation of this instrument should be done by 

Qualified personnel. In order to ensure safe operation, the following instructions should 

be followed.

This instrument has no power-on switch. An external switch or circuit-breaker shall be 

included in the building installation as a disconnecting device.  It shall be marked to 

indicate this function, and it shall be in close proximity to the equipment within easy reach 

of the operator. The switch or circuit-breaker shall not interrupt the Protective Conductor 

(Earth wire), and it shall meet the relevant requirements of IEC  947-1 and IEC 947-3 

(International Electrotechnical Commission). The switch shall not be incorporated in the 

mains supply cord.

Furthermore, to provide protection against excessive energy being drawn from the mains 

supply in case of a fault in the equipment, an over-current protection device shall be 

installed.

• 

The Protective Conductor must be connected for safety reasons. Check that the 

power cable has the proper Earth wire, and it is properly connected. It is not 

safe to operate this unit without the Protective Conductor Terminal connected.

• 

Do not exceed voltage rating on the label located on the top of the 

instrument housing.

• 

Always disconnect power before changing signal and power connections.

• 

Do not use this instrument on a work bench without its case for safety 

reasons.

• 

Do not operate this instrument in flammable or explosive atmospheres.

• 

Do not expose this instrument to rain or moisture.

• 

Unit mounting should allow for adequate ventilation to ensure instrument 

does not exceed operating temperature rating.

• 

Use electrical wires with adequate size to handle mechanical strain and 

power requirements. Install without exposing bare wire outside the connector 

to minimize electrical shock hazards.

EMC Considerations
•  Whenever EMC is an issue, always use shielded cables.
•  Never run signal and power wires in the same conduit.
•  Use signal wire connections with twisted-pair cables.
•  Install Ferrite Bead(s) on signal wires close to the instrument if 

EMC problems persist.

Содержание DP3001

Страница 1: ...http www omega com e mail info omega com User sGuide DP3001 DIGITAL PANEL VOLTMETER ROHS 2 COMPLIANT...

Страница 2: ......

Страница 3: ...nversion 2 2 7 Display 2 2 8 Power 2 2 9 Environmental 2 2 10 Mechanical 3 3 0 MECHANICAL ASSEMBLY INSTALLATION 3 3 1 Case Dimensions 3 3 2 Panel Mounting 4 4 0 POWER SIGNAL 5 4 1 Changing Operating V...

Страница 4: ...1 Side View of Transformer 6 Figure 4 2 Rear Terminal Hookups 7 Figure 5 1 Display Board Jumper Locations 10 Figure 5 2 Main Board Jumper Locations 11 Figure 5 3 Solder Switch Locations 12 Figure 7 1...

Страница 5: ...incorporated in the mains supply cord Furthermore to provide protection against excessive energy being drawn from the mains supply in case of a fault in the equipment an over current protection device...

Страница 6: ......

Страница 7: ...del contains the required circuitry for full range selection and easy configuration 2 0 SPECIFICATIONS 2 1 ANALOG INPUT Model 3 4 5 Range 1 9999 Vdc 19 999 Vdc 199 99 Vdc Resolution 0 1 mV 1 mV 10 mV...

Страница 8: ...ogrammable internally or at the rear connector Overrange indication 4 least significant digits flash 2 8 POWER Standard AC power voltage 115 or 230 Vac 15 AC frequency range 49 to 440 Hz Optional DC p...

Страница 9: ...ouble row of 18 pins 3 96 mm 0 156 in spacing between pins Barrier terminal strip with six 6 screw Bezel Depth behind bezel with connector Panel cutout Weight Case material D1 connector option D4 conn...

Страница 10: ...the panel cutout refer to case dimensions shown below 3 Push the slide retainers onto the case to fit tightly against the panel rear Rotate slide clamps back into original position and tighten clamp s...

Страница 11: ...with the shield terminated to analog ground at the connector Both analog and digital ground are internally connected and should not be connected externally For proper grounding connect the low signal...

Страница 12: ...d Green Green Model Power Signal AC D1 Connector A High 17 High 2 Low 16 Low C Earth Ground T Analog Ground D4 Connector 1 High 6 High 2 Low 5 Low 3 Earth Ground 4 Analog Ground DC D1 Connector 2 DC 1...

Страница 13: ...E24 H 199 99 DP3 7 Spare E23 J 19 999 DP2 8 Spare E22 K 1 9999 DP1 9 Spare E25 L Decimal point select 10 Spare E26 M Spare E27 11 4 6 Vdc power output N Spare E29 12 4 7 Vdc power output P Excitation...

Страница 14: ...s If taken low the converter will complete the full measurement cycle and then hold this reading as long as HOLD is low A positive pulse greater than 300 ns will now initiate a new measurement cycle b...

Страница 15: ...lder switches as indicated Remove all push on jumpers not used 5 1 DECIMAL POINT SELECTION Figure 5 1 Display Board Jumper Locations Decimal Point S1 Alternate decimal point configuration using main b...

Страница 16: ...ons Refer to Figure 5 3 S3 A is used for signal connections Refer to Section 4 for more information Input Configuration Solder Switches Push on Jumpers Wire Jumper Open Close S2 S4 W10 2 Vdc J L H I K...

Страница 17: ...the stability and internal accuracy If calibration is needed use the following procedure 1 Remove the front lens Insert a blade screwdriver under the notch at the bottom of the lens and gently pry it...

Страница 18: ...12 7 0 DRAWINGS Figure 7 1 Assembly Diagram Main and Display Board...

Страница 19: ...t the slope in volts sec during this second period T2 is constant and independent of input signal levels The time required to discharge the capacitor back to zero volts is then proportional to the sig...

Страница 20: ..._______________________ _____________________________________________________________________________ ______________________________________________________________________________ ___________________...

Страница 21: ..._________________________ _____________________________________________________________________________ ______________________________________________________________________________ _________________...

Страница 22: ..._________________________ _____________________________________________________________________________ ______________________________________________________________________________ _________________...

Страница 23: ...NS Equipment sold by OMEGA is not intended to be used nor shall it be used 1 as a Basic Component under 10 CFR 21 NRC used in or with any nuclear installation or activity or 2 in medical applications...

Страница 24: ...Indicators R Turbine Paddlewheel Systems R Totalizers Batch Controllers pH CONDUCTIVITY R pH Electrodes Testers Accessories R Benchtop Laboratory Meters R Controllers Calibrators Simulators Pumps R In...

Отзывы: