GUI Reference
R&S
®
ZNC
153
User Manual 1173.9557.02 ─ 13
"Data to <Destination>" stores the raw trace without the trace functions, "Data & Funct
to <Destination>" stores the trace after it has been transformed using the trace functions.
Remote command:
Show <Destination>
Displays or hides the active memory trace or the memory trace associated with the active
data trace.
●
If no memory trace is associated with the active data trace, "Show <Destination>" is
disabled.
●
If several memory traces are associated with the active data trace, "Show <Destina-
tion>" affects the last generated or changed memory trace.
Remote command:
Show <Data>
Displays or hides the active data trace in the diagram area. If "Trace Math" is active, then
the active mathematical trace is displayed or hidden.
Remote command:
Trace Math
Activates the mathematical mode, applying the last active mathematical relation to the
active data trace. The data trace is replaced by the mathematical trace. "Math" is dis-
played in the trace list while the mathematical mode is active.
"Trace Math" is enabled if the active data trace fulfils the conditions for evaluating the
mathematical relation. E.g. if no "User Defined" mathematical relation is defined, a mem-
ory trace must be coupled to the active data trace, so that the R&S ZNC can evaluate
one of the relations "Data / Mem" or "Data
–
Mem".
Remote command:
Data / Mem, Data
–
Mem
Activates the mathematical mode where the active data trace is divided by (subtracted
from) the last generated memory trace. The division (subtraction) is calculated on a point-
to-point basis: Each measurement point of the active trace is divided by (subtracted from)
the corresponding measurement point of the memory trace. If the memory trace repre-
sents the result of a previous sweep with unchanged settings, the divided (subtracted)
curve is typically centered around 1 / 0 dB (0); it shows the variation of the results in
subsequent sweeps.
The result of the division is a mathematical trace and replaces the active data trace in
the diagram area. The mathematical trace is updated as the measurement goes on and
the analyzer provides new active trace data.
Trace Settings