Command Reference
R&S
®
ZNC
552
User Manual 1173.9557.02 ─ 13
Manual operation:
See
[SENSe<Ch>:]CORRection:COLLect[:ACQuire]:SELected
<Type>, <TestPort>[,
<SecondPort>, <Boolean>, <DelayTimePhase>]
Starts a calibration measurement in order to acquire measurement data for the selected
standards. The standards are reflection or transmission standards and can be connected
to arbitrary analyzer ports.
Note:
The calibration measurement has a variable timeout:
Timeout = (Sweep time /
Number of sweep points) * 3 + 0.1 s
The timeout may be reached e.g. while the measurement waits for the specified trigger
event. It is used for all standard measurements including power calibration sweeps (for
SMARTerCal).
Suffix:
<Ch>
.
Channel number of the calibrated channel
Setting parameters:
<Type>
THRough | OPEN | SHORt | OSHort | OSHort1 | OSHort2 |
OSHort3 | MATCh | NET | ATT | REFL | SLIDe | LINE | LINe1 |
LINe2 | LINe3 | UTHRough | POWer
Standard types: Through, Open, Short, Match, Symmetric Net-
work (NET), Attenuation (ATT), Reflect, Sliding Match (SLIDe),
Line1 (LINE1 and LINE are synonymous), Line2 and Line3 (esp.
for TRL calibration), Offset Short 1 to 3 (OSHort), Unknown
Through, power meter (for SMARTerCal).
<TestPort>
Test port number. For a transmission standard (through, line,
attenuation, symmetric network) the input and output port num-
bers must be specified, for reflection standards, only one port
number is required.
*RST:
n/a
<SecondPort>
Second test port number, for transmission standards only
<Boolean>
Optional status parameter for UTHRough standard:
OFF - unknown through standard is non-dispersive.
ON - unknown through standard is dispersive.
*RST:
OFF
SCPI Command Reference