Command Reference
R&S
®
ZNC
655
User Manual 1173.9557.02 ─ 13
<FirstPort>
Fist port number (sufficient for one-port standards). Port numbers
can be omitted if the cal kit data is valid for all ports; see
<SecondPort>
Second port number, for two-port standards. Port numbers can be
omitted if the cal kit data is valid for all ports.
Example:
See
Usage:
Query only
Manual operation:
See
SYSTem:COMMunicate:RDEVice:AKAL:WARMup[:STATe]?
Queries the warmup state of the connected calibration unit R&S ZV_Z5x. If several cal
units are USB-connected to the analyzer, the command queries the cal unit selected via
.
Possible responses are
1
(true, if the calibration unit has been connected for a sufficient
time to rach its operating temperature) or
0
(false).
0
is also returned if no calibration unit
is connected.
Example:
See
Usage:
Query only
Manual operation:
See
SYSTem:COMMunicate:RDEVice:PMETer<Pmtr>:AZERo
Starts auto zeroing of the external power meter.
Suffix:
<Pmtr>
.
Number of the configured power meter. Power meters must be
numbered in ascending order, starting with 1. If a number is re-
used, the previous power meter configuration is overwritten.
Power meters can be assigned several times so that the number
of configured power meters is practically unlimited.
Example:
See
Usage:
Event
Manual operation:
See
SYSTem:COMMunicate:RDEVice:PMETer:DELete
Clears the configuration table for external power meters.
SCPI Command Reference