Command Reference
R&S
®
ZNC
540
User Manual 1173.9557.02 ─ 13
Parameters:
<CalKitName>
String parameter containing the name of a calibration kit available
on the analyzer. See "Cal kit naming conventions " above.
*RST:
n/a (A *RST does not change the assignment
between connector types and calibration kits.)
Example:
MMEM:LOAD:CKIT 'C:\Users\Public
\Documents\Rohde-Schwarz
\Vna\Calibration\Kits\New Kit 1.calkit'
Load the previously created cal kit file
New Kit 1.calkit
from
the default cal kit directory.
CORR:CKIT:N50:SEL 'New Kit 1'
Assign the imported kit to the N 50 Ω connector type, assuming
that the cal kit name stored in
New Kit 1.calkit
reads
New
Kit 1
.
Manual operation:
See
"Add / Copy / Delete / Standards..."
[SENSe:]CORRection:CKIT:CATalog?
<ConnectorType>
Returns a list of all cal kits for a given connector type.
Parameters:
<ConnectorType>
Name of the connector type. Use
names.
Example:
See
Usage:
Query only
Manual operation:
See
"Add / Copy / Delete / Standards..."
[SENSe:]CORRection:CKIT:DELete
<CalKitName>
Deletes an imported or user-defined cal kit.
Note:
It is not possible to modify or store predefined or ideal kits.
Setting parameters:
<CalKitName>
String parameter containing an imported or user-defined calibra-
tion kit available on the analyzer.
Example:
See
Usage:
Setting only
Manual operation:
See
"Add / Copy / Delete / Standards..."
[SENSe:]CORRection:CKIT:LABel
<CalKitName>[, <Label>]
Assigns a label to an imported or user-defined calibration kit.
SCPI Command Reference