Command Reference
R&S
®
ZNC
437
User Manual 1173.9557.02 ─ 13
Manual operation:
See
CALCulate<Chn>:RIPPle:SOUNd[:STATe]
<Boolean>
Switches the acoustic signal (fail beep) on or off. The fail beep is generated each time
the analyzer detects an exceeded ripple limit.
Suffix:
<Chn>
.
Channel number used to identify the active trace.
Parameters:
<Boolean>
ON | OFF - fail beep on or off.
*RST:
OFF
Example:
CALC:RIPP:STAT ON; SOUN ON
Switch the limit check on and activate the fail beep.
Manual operation:
See
CALCulate<Chn>:RIPPle:STATe
<Boolean>
Switches the ripple limit check for the active trace on or off.
Tip:
Use
check for a single ripple limit range on or off.
Suffix:
<Chn>
.
Channel number used to identify the active trace.
Parameters:
<Boolean>
ON | OFF
–
ripple limit check on or off.
*RST:
OFF
Example:
See
Manual operation:
See
CALCulate<Chn>:RIPPle:STATe:AREA
<HorizontalPos>, <VerticalPos>
Moves the ripple test info field for the active trace <Chn> to one of nine predefined posi-
tions in the active diagram.
Suffix:
<Chn>
.
Channel number used to identify the active trace.
Parameters:
<HorizontalPos>
LEFT | MID | RIGHt
Horizontal position
<VerticalPos>
TOP | MID | BOTTom
Vertical position
Example:
See
SCPI Command Reference