GUI Reference
R&S
®
ZNC
187
User Manual 1173.9557.02 ─ 13
Defining ripple limits with minimum effort
Choose one of the following methods to efficiently create and handle ripple limit ranges:
●
To configure a limit test with only a few ranges, use "Add" and edit each range in the
table individually.
●
Use the "Align All" button to create non-overlapping, contiguous ranges of equal
width.
●
Save your ripple ranges to a file so you can re-use or modify them later sessions
("Save Ripple Test..., Recall Ripple Test...").
Background information
Refer to
Access:
TRACE > LINES > Ripple Test > Def. Ripple Test...
The "Define Ripple Test" dialog contains a table to edit the individual ranges of the ripple
check ranges. The buttons below the table extend, shorten, or re-order the range list and
save/recall ripple test data.
Range List
Defines the individual ripple limit ranges.
The table contains an automatically assigned current number for each range plus the
following editable columns:
●
"Range On/Off" enables or disables the ripple limit check in each range. Disabling
the ripple limit check does not delete the range but hides the entry in the info field.
●
"Start Stimulus" is the smallest stimulus (x-axis) value of the range.
●
"Stop Stimulus" is the largest stimulus (x-axis) value of the range.
●
"Ripple Limit" is the maximum allowed difference between the largest and the small-
est trace value in the range.
Trace Settings