Command Reference
R&S
®
ZNC
691
User Manual 1173.9557.02 ─ 13
Example:
CORR:CKIT:N50:FOPEN 'ZV-Z21','',
0,1.8E+010,0.0151,0,0,0.22,-0.22,0.0022
Define the properties of the open (f) standard for the N 50 Ω con-
nector type contained in the ZV-Z21 calibration kit: Assign a valid
frequency range of 0 Hz to 18 GHz, an electrical length of 15.1
mm, 0 dB loss and define the polynomial coefficients of the fringing
capacitance as 0 fF, 0.22 fF/GHz, –0.22 fF/(GHz)
2
, 0.0022 fF/
(GHz)
3
.
[SENSe:]CORRection:CKIT:INSTall
<CalKitFile>
Loads cal kit data from a specified R&S ZVR cal kit file.
Setting parameters:
<CalKitFile>
String parameter to specify the name and directory of the cal kit
file to be loaded.
Note:
The loaded file must be a R&S ZVR-specific cal kit file with
the extension
*.ck
. VNA cal kit files (
*.calkit
) can be imported
using the
command. Agilent cal kit files
can be imported manually and converted into
*.calkit
files.
Example:
CORR:CKIT:INST 'C:\Users\Public\Documents
\Rohde-Schwarz\Vna\Calibration\Kits\ZCAN.ck'
Load the previously created R&S ZVR cal kit file
ZCAN.ck
from
the default cal kit directory.
MMEM:STOR:CKIT 'ZCAN', 'C:
\Users\Public\Documents
\Rohde-Schwarz\Vna\Calibration\Kits\ZCAN.calkit'
Store the imported cal kit data to a VNA cal kit file
ZCAN.calkit
(assuming that the cal kit name stored in
ZCAN.ck
reads "ZCAN").
Usage:
Setting only
[SENSe<Ch>:]CORRection:COLLect[:ACQuire]
<Standard>[, <Dispersion>,
<Delay>]
Starts a calibration measurement in order to acquire measurement data for the selected
standards. The standards are reflection or transmission standards and must be connec-
ted to port 1 or 2 of the analyzer.
Tip:
to obtain measurement data at arbitrary analyzer ports.
Suffix:
<Ch>
.
Channel number of the calibrated channel.
R&S ZVR/ZVAB Compatible Commands