Remote Commands in the Receiver Application
R&S
®
ESW
636
User Manual 1177.6298.02 ─ 08
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
[SENSe:]CORRection:TRANsducer:CATalog?
This command queries all transducer factors stored on the R&S
ESW.
Parameters:
<FileInformation>
String containing the file information.
'<NameFileN>,<SizeFileN>'
<NameFileN>
Name of the file.
<SizeFileN>
Size of the file in bytes.
Return values:
<UsedDiskSpace>
Size of all available files containing transducer factors in byte.
<FreeDiskSpace>
Remaining disk space in bytes.
Example:
CORR:TRAN:CAT?
would return, e.g.
1579,34482896896,'FACTOR1,823','FACTOR2,756'
Usage:
Query only
[SENSe:]CORRection:TRANsducer:COMMent
<Comment>
This command defines the comment for the selected transducer factor.
Before you can use the command, you have to select and turn on a transducer.
Parameters:
<Comment>
*RST:
(empty comment)
[SENSe:]CORRection:TRANsducer:DATA
{<Frequency>, <Level>}...
This command configures transducer factors for specific trace points. A set of trans-
ducer factors defines an interpolated transducer line and can be stored on the instru-
ment.
Parameters:
<Frequency>
The unit for <Frequency> is Hz, which may or may not be omit-
ted. Frequencies have to be sorted in ascending order.
Default unit: Hz
<Level>
The unit for <Level> depends on
General Instrument Setup