Remote Commands in the Receiver Application
R&S
®
ESW
662
User Manual 1177.6298.02 ─ 08
You can define the message with
Parameters:
<State>
ON
Displays a custom message.
Shows a message "Remote Mode controlled by <string>",
including the network address of the controlling device.
OFF
Displays the standard message ("Remote mode").
*RST:
OFF
Example:
//Define and display custom message
SYST:DISP:MESS 'a computer'
SYST:DISP:MESS:STAT ON
//would display:
Remote mode controlled by a computer
SYSTem:DISPlay:UPDate
<State>
This command turns the display during remote operation on and off.
If on, the R&S
ESW updates the diagrams, traces and display fields only.
The best performance is obtained if the display is off during remote control operation.
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
Example:
SYST:DISP:UPD ON
Manual operation:
See
SYSTem:ERRor:CLEar:REMote
This command deletes all contents of the "Remote Errors" table.
Note: The remote error list is automatically cleared when the R&S
ESW is shut down.
Example:
SYST:ERR:CLE:REM
Manual operation:
See
See
SYSTem:ERRor:DISPlay
<State>
This command the error display during remote operation on and off.
If activated, the R&S
ESW displays a message box at the bottom of the screen that
contains the most recent type of error and the command that caused the error.
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
General Instrument Setup