Remote Commands in the Receiver Application
R&S
®
ESW
550
User Manual 1177.6298.02 ─ 08
If the marker is not yet active, the command first activates the marker.
Suffix:
<n>
.
<m>
Manual operation:
See
CALCulate<n>:DELTamarker<m>:MINimum:RIGHt
This command moves a delta marker to the next higher minimum value.
The search includes only measurement values to the right of the current marker posi-
tion.
In the spectrogram, the command moves a marker horizontally to the minimum level in
the currently selected frame. The vertical marker position remains the same.
Suffix:
<n>
.
<m>
Manual operation:
See
15.7.4.5
Marker Results
Commands to retrieve the marker position described elsewhere.
●
●
CALCulate<n>:DELTamarker<m>:X:RELative?
................................................................. 550
................................................................................. 551
........................................................................................ 551
CALCulate<n>:DELTamarker<m>:X:RELative?
This command queries the relative position of a delta marker on the x-axis.
If necessary, the command activates the delta marker first.
Suffix:
<n>
.
<m>
Return values:
<Position>
Position of the delta marker in relation to the reference marker.
Example:
CALC:DELT3:X:REL?
Outputs the frequency of delta marker 3 relative to marker 1 or
relative to the reference position.
Usage:
Query only
Analysis