Common Measurement Settings
R&S
®
ESW
152
User Manual 1177.6298.02 ─ 08
TTL synchronization
In addition, TTL synchronization can be used with some Rohde
&
Schwarz generators
connected via GPIB. The TTL interface is included in the AUX control connector of the
External Generator Control option.
Using the TTL interface allows for considerably higher measurement rates than pure
GPIB control, because the frequency stepping of the R&S
ESW is directly coupled with
the frequency stepping of the generator. For details see
In
the TTL connection is illustrated using an R&S SMU generator, for
example.
R&S SMU rear
Analyzer rear
BNC Trigger
BNC Blank
Figure 10-1: TTL connection for an R&S SMU generator
In
, the connection for an R&S SMW is shown.
Signal generator
rear panel
Signal analyzer
rear panel
BNC Blank
BNC Trigger
Figure 10-2: TTL connection for an R&S SMW generator
The external generator can be used to calibrate the data source by performing either
transmission or reflection measurements.
Transmission Measurement
This measurement yields the transmission characteristics of a two-port network. The
external generator is used as a signal source. It is connected to the input connector of
the DUT. The input of the R&S
ESW is fed from the output of the DUT. A calibration
Data Input and Output