Remote Commands in the Receiver Application
R&S
®
ESW
627
User Manual 1177.6298.02 ─ 08
DIAGnostic:SERVice:INPut:PULSed:CFRequency
<Frequency>
This command defines the frequency of the calibration signal.
Before you can use the command, you have to feed in a calibration signal with
DIAGnostic:SERVice:INPut[:SELect]
Parameters:
<Frequency>
Possible frequencies of the calibration signal are fix.
If you define a frequency that is not available, the R&S
ESW
uses the next available frequency. Example: a frequency of
20
MHz is rounded up to the next available frequency (25
MHz).
*RST:
50 MHz
Default unit: Hz
Manual operation:
See
DIAGnostic:SERVice:INPut:RF[:SPECtrum]
<Bandwidth>
This command selects the bandwidth of the calibration signal.
Parameters:
<Bandwidth>
NARRowband | BROadband
NARRowband
Narrowband signal for power calibration of the frontend.
BROadband
Broadband signal for calibration of the IF filter.
Manual operation:
See
DIAGnostic:SERVice:INPut[:SELect]
<Signal>
This command activates or deactivates the use of an internal calibration signal as input
for the R&S
ESW.
Parameters:
<Signal>
CALibration
Uses the calibration signal as RF input.
MCALibration
Uses the calibration signal for the microwave range as RF input.
RF
Uses the signal from the RF input.
*RST:
RF
Example:
DIAG:SERV:INP CAL
Uses the calibration signal as RF input.
Manual operation:
See
See
See
General Instrument Setup