Remote Commands in the Receiver Application
R&S
®
ESW
658
User Manual 1177.6298.02 ─ 08
●
INSTrument:COUPle:USER<j>:REMove
●
INSTrument:COUPle:USER<j>:RELation
Suffix:
<j>
.
irrelevant
Return values:
<Index>
Index numbers of all available custom couplings (as a comma-
separated list of strings).
Note that the returned index numbers do not necessarily have to
be the same as those shown in the user interface.
Example:
INST:COUP:USER:NUMB:LIST?
would return, e.g.:
'1','2','4'
(Number '3' is not returned, because a coupling with that index
does not exist anymore.)
Usage:
Query only
INSTrument:COUPle:USER<j>:RELation
<Direction>
This command selects the direction in which synchronization works.
Note that the command is not available if you synchronize over all channels or all
channels of the same application.
Suffix:
<j>
.
Selects the index of a custom synchronization mechanism.
You can query the index with
Parameters:
<Direction>
BIDir
Changes of a parameter get applied both ways (from channel 1
to channel 2 and vice versa).
LTOR
Changes of a parameter get applied from channel 1 to channel
2, but not the other way around.
RTOL
Changes of a parameter get applied from channel 2 to channel
1, but not the other way around.
Example:
INST:COUP:USER:REL BID
Selects bidirectional changes for the custom coupling with the
index number 1.
INSTrument:COUPle:USER<j>:REMove
[<Scope>]
This command deletes a custom coupling mechanism.
General Instrument Setup