Data Management
R&S
®
ESW
265
User Manual 1177.6298.02 ─ 08
Access
: "Save"/ "Open" icon in the toolbar > "Save" / "Recall"
Both dialog boxes are very similar and closely related.
Selecting Storage Location - Drive/ Path/ Files
...........................................................265
Recall in New Channel / Recall in Current Channel
....................................................267
Selecting Storage Location - Drive/ Path/ Files
Select the storage location of the file on the <instrument> or an external drive.
Note:
Saving instrument settings in secure user mode.
In secure user mode, settings that are stored on the instrument are stored to volatile
memory, which is restricted to 256
MB. Thus, a "memory limit reached" error can occur
although the hard disk indicates that storage space is still available.
To store data permanently, select an external storage location such as a USB memory
device.
For details, see
Chapter 4.7, "Protecting Data Using the Secure User Mode"
Remote command:
File Name
Contains the name of the data file without the path or extension.
File names must be compatible with the Windows conventions for file names. In partic-
ular, they must not contain special characters such as ":", "*", "?".
Storing and Recalling Instrument Settings and Measurement Data