General Instrument Setup
R&S
®
ESW
322
User Manual 1177.6298.02 ─ 08
●
"Name" / "Comment"
Defines a name and / or comment of the transducer set.
●
"Unit"
Selects the unit of the transducer set.
●
"Break"
Turns the transducer break on and off.
●
"Insert Range"
Adds a new range to the transducer set. Each transducer set can consist of up to
10 ranges.
To select a range, select the corresponding range in the table, or select it with the
buttons on the bottom of the dialog box.
●
"Delete Range"
Deletes the currently selected transducer set range. The currently selected range
is highlighted orange.
●
"Start Frequency" / "Stop Frequency"
Defines the start and stop frequencies for the selected transducer set range.
●
"Add Factor"
Opens a list of transducer factors that you can apply to the transducer set range.
You can assign up to eight transducer factors to each range, but only those whose
characteristics (frequency range, unit) are compatible to the characteristics of the
selected range.
●
"Change Factor"
Replaces the selected transducer factor with another one. The selected transducer
factor is highlighted.
●
"Remove Factor"
Removes the selected transducer factor from the transducer set range.
●
"Save Sets"
Saves and stores the transducer set on the internal hard disk of the R&S
ESW.
In secure user mode, settings that are stored on the instrument are stored to volatile
memory, which is restricted to 256
MB. Thus, a "memory limit reached" error can occur
although the hard disk indicates that storage space is still available.
To store data permanently, select an external storage location such as a USB memory
device.
For details, see
Chapter 4.7, "Protecting Data Using the Secure User Mode"
Remote command:
See
Transducers