![Rohde & Schwarz ESW Series User Manual Download Page 513](http://html.mh-extra.com/html/rohde-and-schwarz/esw-series/esw-series_user-manual_1477816513.webp)
Remote Commands in the Receiver Application
R&S
®
ESW
513
User Manual 1177.6298.02 ─ 08
*RST:
NORMal
Example:
BAND:TYPE NORM
Manual operation:
See
15.6.7
Trigger Configuration
....................................................................................513
.....................................................................................513
....................................................................................513
................................................................................... 514
........................................................................................515
TRIGger<tp>[:SEQuence]:HOLDoff[:TIME]
....................................................................... 515
TRIGger<tp>[:SEQuence]:LEVel[:EXTernal]
......................................................................515
....................................................................................516
................................................................................. 516
[SENSe:]SWEep:EGATe:HOLDoff
<DelayTime>
This command defines the delay time between the gate signal and the continuation of
the measurement.
Parameters:
<DelayTime>
Range:
0 s to 30 s
*RST:
0 s
Default unit: S
Example:
SWE:EGAT:HOLD 100us
Manual operation:
See
[SENSe:]SWEep:EGATe:LENGth
<GateLength>
This command defines the gate length.
Parameters:
<GateLength>
Range:
125 ns to 30 s
*RST:
400μs
Default unit: S
Example:
SWE:EGAT:LENG 10ms
Manual operation:
See
[SENSe:]SWEep:EGATe:POLarity
<Polarity>
This command selects the polarity of an external gate signal.
The setting applies both to the edge of an edge-triggered signal and the level of a
level-triggered signal.
Configuration