Remote Commands in the Receiver Application
R&S
®
ESW
507
User Manual 1177.6298.02 ─ 08
CALCulate<n>:UNIT:POWer
<Unit>
This command selects the unit for the y-axis.
Suffix:
<n>
.
irrelevant
Parameters:
<Unit>
DBM | V | A | W | DBPW | DBPW_MHZ | DBUV | DBUV_MHZ |
DBMV | DBMV_MHZ | DBUA | DBUA_MHZ | DBPT |
DBPT_MHZ | DBUV_M | DBUV_MMHZ | DBUA_M |
DBUA_MMHZ
*RST:
dBµV
Example:
//Select unit
CALC:UNIT:POW DBM
Manual operation:
See
See
INPut<ip>:ATTenuation[:VALue]
<Attenuation>
This command defines the attenuation at the RF input.
To protect the input mixer, attenuation levels of 10
dB or less are possible only if you
have turned off the input protection with
INPut<ip>:ATTenuation:PROTection[:
Suffix:
<ip>
.
irrelevant
Parameters:
<Attenuation>
Range:
0 dB to 79 dB
*RST:
10 dB
Default unit: dB
Example:
//Define attenuation
INP:ATT 40dB
Manual operation:
See
INPut<ip>:ATTenuation:AMODe
<State>
This command selects the auto ranging mode.
Suffix:
<ip>
.
irrelevant
Parameters:
<State>
LOWNoise
Selects the low noise mode.
NORMal
Selects the normal mode.
*RST:
NORMal
Configuration