![Rohde & Schwarz ESW Series User Manual Download Page 457](http://html.mh-extra.com/html/rohde-and-schwarz/esw-series/esw-series_user-manual_1477816457.webp)
Remote Commands in the Receiver Application
R&S
®
ESW
457
User Manual 1177.6298.02 ─ 08
[SENSe:]SCAN<sr>:INPut:ATTenuation:AUTO
<State>
This command turns auto ranging in a scan range on and off.
Suffix:
<sr>
.
irrelevant
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
OFF
Example:
//Turn off auto ranging in the 4th scan range
SCAN4:INP:ATT:AUTO OFF
Manual operation:
See
[SENSe:]SCAN<sr>:INPut:ATTenuation[:VALue]
<Attenuation>
This command defines the attenuation level applied in the selected scan range.
Suffix:
<sr>
.
Selects the scan range.
Parameters:
<Attenuation>
<numeric value> (integer only)
Range:
0 dB to 79 dB
Increment: 1 dB
*RST:
10 dB
Default unit: dB
Example:
//Define attenuation for the 4th scan range.
SCAN4:INP:ATT 30dB
Manual operation:
See
[SENSe:]SCAN<sr>:INPut:GAIN:AUTO
<State>
This command includes and excludes the preamplifier from the auto ranging feature.
Suffix:
<sr>
.
Selects the scan range.
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
OFF
Example:
//Automatically determine if the preamplifier is used or not
SCAN1:INP:GAIN:AUTO ON
Manual operation:
See
Measurements and Result Displays