Remote Commands in the Receiver Application
R&S
®
ESW
663
User Manual 1177.6298.02 ─ 08
Example:
SYST:ERR:DISP ON
Manual operation:
See
SYSTem:HPCoupling
<CouplingType>
Controls the default coupling ratios in the HP emulation mode for:
●
span and resolution bandwidth (Span/RBW) and
●
resolution bandwidth and video bandwidth (RBW/VBW)
This command is only available if a HP language is selected using
Parameters:
<CouplingType>
HP | FSP
*RST:
FSP
Example:
SYSTem:HPC HP
Manual operation:
See
SYSTem:IFGain:MODE
<Mode>
Configures the internal IF gain settings in HP emulation mode due to the application
needs. This setting is only taken into account for resolution bandwidth < 300 kHz and is
only available if a HP language is selected using
Parameters:
<Mode>
NORMal | PULSe
NORMal
Optimized for high dynamic range, overload limit is close to ref-
erence level.
PULSe
Optimized for pulsed signals, overload limit up to 10 dB above
reference level.
*RST:
NORM
Example:
SYST:IFG:MODE PULS
Manual operation:
See
SYSTem:IDENtify:FACTory
This command resets the query to
to its default value.
Manual operation:
See
SYSTem:IDENtify[:STRing]
<String>
This command defines the response to
General Instrument Setup