Remote Commands in the Receiver Application
R&S
®
ESW
530
User Manual 1177.6298.02 ─ 08
CALCulate<n>:MATH<t>:MODE
<Mode>
This command selects the way the R&S
ESW calculates trace mathematics.
Suffix:
<n>
.
<t>
irrelevant
Parameters:
<Mode>
For more information on the way each mode works see
LINear
Linear calculation.
LOGarithmic
Logarithmic calculation.
POWer
Linear power calculation.
*RST:
LOGarithmic
Example:
CALC:MATH:MODE LIN
Selects linear calculation.
Manual operation:
See
CALCulate<n>:MATH<t>:POSition
<Position>
This command defines the position of the trace resulting from the mathematical opera-
tion.
Suffix:
<n>
.
<t>
irrelevant
Parameters:
<Position>
Vertical position of the trace in % of the height of the diagram
area.
100 PCT corresponds to the upper diagram border.
Range:
-100 to 200
*RST:
50
Default unit: PCT
Example:
CALC:MATH:POS 100
Moves the trace to the top of the diagram area.
Manual operation:
See
CALCulate<n>:MATH<t>:STATe
<State>
This command turns the trace mathematics on and off.
Suffix:
<n>
.
Analysis