Common Analysis and Display Functions
R&S
®
ESW
252
User Manual 1177.6298.02 ─ 08
"File Explorer": Instead of using the file manager of the R&S
ESW firmware, you can
also use the Microsoft Windows File Explorer to manage files.
The limit line can be imported again later by the R&S
ESW for use in other measure-
ments.
Remote command:
File Explorer ← Export
Opens the Microsoft Windows File Explorer.
Remote command:
not supported
11.5.2.3
How to Define Limit Lines
Access
: "Overview" > "Analysis" > "Lines" > "Limit Lines"
The following tasks are described here:
●
"How to find compatible limit lines"
●
"How to activate and deactivate a limit check"
●
"How to edit existing limit lines"
●
"How to copy an existing limit line"
●
"How to delete an existing limit line"
●
"How to configure a new limit line"
●
"How to move the limit line vertically or horizontally"
How to find compatible limit lines
► In the "Line Config" dialog box, select the "View Filter" option: "Show Compatible".
All stored limit lines with the file extension
.LIN
in the
limits
subfolder of the
main installation folder of the instrument that are compatible to the current mea-
surement settings are displayed in the overview.
How to activate and deactivate a limit check
A limit check is automatically performed as soon as any of the limit lines is activated.
1. To activate a limit check:
Select the "Check Traces" setting for a limit line in the overview and select the
trace numbers to be included in the limit check. One limit line can be assigned to
several traces.
The specified traces to be checked are compared with the active limit lines. The
status of the limit check is indicated in the diagram.
2. To deactivate a limit line, deactivate all "Traces to be Checked" for it.
To deactivate all limit lines at once, select the "Disable All Lines" button.
The limit checks for the deactivated limit lines are stopped and the results are
removed form the display.