Common Analysis and Display Functions
R&S
®
ESW
238
User Manual 1177.6298.02 ─ 08
Functions to position markers in the Spectrum application described elsewhere:
●
.................................................................................................238
Peak Search
Sets the selected marker/delta marker to the maximum of the trace. If no marker is
active, marker 1 is activated.
For spectrogram displays, define which frame the peak is to be searched in.
Remote command:
CALCulate<n>:MARKer<m>:MAXimum[:PEAK]
CALCulate<n>:DELTamarker<m>:MAXimum[:PEAK]
Search Next Peak
Sets the selected marker/delta marker to the next (lower) maximum of the assigned
trace. If no marker is active, marker 1 is activated.
For spectrogram displays, define which frame the next peak is to be searched in.
Remote command:
CALCulate<n>:MARKer<m>:MAXimum:NEXT
CALCulate<n>:MARKer<m>:MAXimum:RIGHt
CALCulate<n>:MARKer<m>:MAXimum:LEFT
CALCulate<n>:DELTamarker<m>:MAXimum:NEXT
CALCulate<n>:DELTamarker<m>:MAXimum:RIGHt
CALCulate<n>:DELTamarker<m>:MAXimum:LEFT
Search Minimum
Sets the selected marker/delta marker to the minimum of the trace. If no marker is
active, marker 1 is activated.
For spectrogram displays, define which frame the minimum is to be searched in.
Remote command:
CALCulate<n>:MARKer<m>:MINimum[:PEAK]
CALCulate<n>:DELTamarker<m>:MINimum[:PEAK]
Search Next Minimum
Sets the selected marker/delta marker to the next (higher) minimum of the selected
trace. If no marker is active, marker 1 is activated.
For spectrogram displays, define which frame the next minimum is to be searched in.
Remote command:
CALCulate<n>:MARKer<m>:MINimum:NEXT
CALCulate<n>:MARKer<m>:MINimum:LEFT
CALCulate<n>:MARKer<m>:MINimum:RIGHt
CALCulate<n>:DELTamarker<m>:MINimum:NEXT
CALCulate<n>:DELTamarker<m>:MINimum:LEFT
CALCulate<n>:DELTamarker<m>:MINimum:RIGHt
Marker Usage