Remote Commands in the Receiver Application
R&S
®
ESW
445
User Manual 1177.6298.02 ─ 08
A detailed programming example is provided in
Chapter 15.5.13, "Programming Exam-
ple: Performing a Sequence of Measurements"
Parameters:
<State>
ON | OFF | 0 | 1
ON | 1
The Sequencer is activated and a sequential measurement is
started immediately.
OFF | 0
The Sequencer is deactivated. Any running sequential measure-
ments are stopped. Further Sequencer commands
(
INIT:SEQ...
) are not available.
*RST:
0
Example:
SYST:SEQ ON
Activates the Sequencer.
INIT:SEQ:MODE SING
Sets single Sequencer mode so each active measurement will
be performed once.
INIT:SEQ:IMM
Starts the sequential measurements.
SYST:SEQ OFF
Manual operation:
See
15.5.3
Result Retrieval
............................................................................... 445
15.5.3.1
Trace Data and Result Query
This section contains information on the
TRACe:DATA
command and a detailed
description of the characteristics of that command. Basically, the command queries the
results of the current measurement. The command supports various SCPI parameters
in combination with the query. Each SCPI parameter returns a different aspect of the
measurement.
The format of the return values is either in ASCII or binary characters and depends on
the format you have set with
Querying trace data
The SCPI parameters
TRACE1
| ... |
TRACE6
return the trace data for the correspond-
ing trace.
Example:
TRAC? TRACE1
Measurements and Result Displays