Remote Commands in the Receiver Application
R&S
®
ESW
461
User Manual 1177.6298.02 ─ 08
[SENSe:]SCAN<sr>:STOP
<Frequency>
This command defines the stop frequency of a scan range.
Suffix:
<sr>
.
Selects the scan range.
Parameters:
<Frequency>
<numeric value>
Range:
Refer to the datasheet
*RST:
Depends on the scan range.
Default unit: Hz
Example:
//Define stop frequency for 4th scan range
SCAN4:STOP 500MHz
Manual operation:
See
[SENSe:]SCAN<sr>:TIME
<Time>
This command defines the measurement time applied in the selected scan range.
Suffix:
<sr>
.
Selects the scan range.
Parameters:
<Time>
<numeric value>
Range:
100 µs to 100 s
*RST:
depends on the range
Default unit: s
Example:
//Define measurement time for 4th scan range
SCAN4:TIME 1 ms
Manual operation:
See
[SENSe:]SWEep:COUNt
<Measurements>
This command defines the number of measurements that the application uses to aver-
age traces.
For continuous measurements, the application calculates the moving average over the
average count.
For single measurements, the application stops the measurement and calculates the
average after the average count has been reached.
Measurements and Result Displays