Remote Commands in the Receiver Application
R&S
®
ESW
557
User Manual 1177.6298.02 ─ 08
Parameters:
<Frame>
Selects a frame either by its frame number or time stamp.
The frame number is available if the time stamp is off. The range
depends on the history depth.
The time stamp is available if the time stamp is on. The number
is the distance to frame 0 in seconds. The range depends on the
history depth.
Default unit: S
Example:
CALC:DELT4:SGR:FRAM -20
Sets fourth deltamarker 20 frames below marker 1.
CALC:DELT4:SGR:FRAM 2 s
Sets fourth deltamarker 2 seconds above the position of marker
1.
Manual operation:
See
CALCulate<n>:DELTamarker<m>:SGRam:SARea
<SearchArea>
CALCulate<n>:DELTamarker<m>:SPECtrogram:SARea
<SearchArea>
This command defines the marker search area for
all
spectrogram markers in the
channel.
Parameters:
<SearchArea>
VISible
Performs a search within the visible frames.
Note that the command does not work if the spectrogram is not
visible for any reason (e.g. if the display update is off).
MEMory
Performs a search within all frames in the memory.
*RST:
VISible
Manual operation:
See
CALCulate<n>:DELTamarker<m>:SGRam:XY:MAXimum[:PEAK]
CALCulate<n>:DELTamarker<m>:SPECtrogram:XY:MAXimum[:PEAK]
This command moves a marker to the highest level of the spectrogram over all fre-
quencies.
Suffix:
<n>
.
<m>
CALCulate<n>:DELTamarker<m>:SGRam:XY:MINimum[:PEAK]
CALCulate<n>:DELTamarker<m>:SPECtrogram:XY:MINimum[:PEAK]
This command moves a delta marker to the minimum level of the spectrogram over all
frequencies.
Analysis